| Specifications |
| Measurement Ranges: |
| CO Ranges: |
0 - 50 ppm up to 0 - 2 000 ppm |
| Resolution: |
1 ppm / 1 mg/m3 |
| Linearity: |
2% of range |
| Zero drift: |
2% of range per month |
| Span drift: |
2% of range per month |
| 02 Ranges: |
0 - 5 % to 0 - 25% |
| Resolution: |
0.1 Vol % |
| Linearity: |
0.2 Vol % |
| Zero drift: |
0.2 Vol % per month |
| Span drift: |
0.2 Vol % per month |
| Calibration: |
Microprocessor controlled |
| Standard: |
Manual gas selection |
| Display: |
LCD with backlight |
| Indicators |
External "power On" and "System OK" LEDs |
| Outputs/Inputs |
| Analog outputs: |
Isolated current loop outputs
Efficiency (on FGA 930E only)
0, 2, or 4 to 10 or 20 mA |
| Relay outputs: |
System OK; Maintenance/Calibration |
| Relay rating: |
Isolated changeover S.P - 1 A @ 240 V a.c.
5 A @ 240 V d.c. resistive for Alarm - one per gas measured
Span and Zero Gas Relays for calibration gases |
| Inputs: |
Current loop inputs for ambient and process temperatures (only needed for FGA 930 E) |
| Environmental |
|
| Environmental rating: |
IP65 / NEMA 4 |
| Operating (ambient) temp: |
32 to 95ºF / 0 to +35ºC standard to -4ºF / -20ºC with optional case heater to 122ºF / +50ºC with optional air conditioner |
| Compliance |
| Measurement standards |
Meets the requirements of ISO 12039, ISO 10849 and ISO 7935 |
| Approvals: |
MCerts for Continuous Emissions Monitoring Systems (Nov 1988)
TUV approved to the 13th & 17th BlmSchV and TA Luft USEPA certified to PS-3, PS-4 and PS-4A |
| Electrical Safety: |
Conforms to EN-61010-2 |
| EMC: |
Conforms to EN-50 081 & EN-50 082 |
| Power |
| Power Supply: |
110 V a.c. or 230 V a.c., +20 %, 50 - 60 Hz |
| Power consumption: |
300 W |
| Gas and Air Requirements |
| Instrument air (zero calibration): |
2 bar / 30 psi clean and dry, 5 l/min / 0.2 cfm |
| Instrument air (cooling): |
2 - 10 bar / 30 - 150 psi clean and dry, 90 l/min / 3 cfm |
| Calibration gas (recommended): |
2 bar / 30 psi, 5 l/min / 0.2 cfm 20 litres / 0.7 cu. ft. per calibration approx. |
| Calibration gas type: |
Specific to each gas type and measurement span |
| Dimensions (HxWxD): |
24 x 24 x 14 in / 800 x 350 mm |
| Weight: |
117 lb / 53 kg |
| Options |
Special measurement ranges
Automatic calibration
Efficiency measurement (FGA 930 E only)
Probe type
Sample line type
Case heater
Air conditioner/vortex cooler |