 Critical moisture measurement applications include high purity gas production; olefin manufacture, storage, and transmission; semiconductor gases; and industrial gas production and quality assurance. For these applications and more, turn to AMETEK Process Instruments’ Model 5800 Moisture Analyzer. The Model 5800 offers a truly remarkable combination of performance benefits: exceptional accuracy, multi-gas compatibility, fast response speed, and wide measurement range.
Model 5800 Specifications
| Compatible Gases: |
Inerts (He, Ar, Ne, Xe, Kr), O2 , H2, N2, NO, CO, hydrocarbons, air, and some specialty gases such as sulfur hexafluoride, CO2, requires a custome measurement cell. Contact AMETEK to confirm compatibility with other gases. |
| Range: |
0.02 to 100 ppmv. Indicates trend to 1000 ppmv. Display is software settable to show ppmv, ppbv or dew point (requires pressure input) |
| Limit of Detection: |
0.02 ppm |
| Accuracy: |
±20 ppbv or 5% of the reading, whichever is greater |
| Sensitivity: |
5 ppbv or 0.5% of instrument range, whichever is greater |
| Response Time: |
63% of a step change in either direction in less than 5 minutes |
| Inlet Pressure: |
1 to 6.89 Bar (15 to 100 psig) |
| Exhaust Pressure: |
Atmospheric |
| Sample Flow Requirements: |
600 ml/min at STP |
| Sample Gas Temperature: |
0º to 100ºC (32º to 212ºF) |
| Outputs: |
Four line x 20 character vacuum fluorescent digital display
Two fully programmable 4 to 20 mA analog outputs, into 1200 ohm load
RS485 bidirectional serial port |
| Alarms: |
Three independent contact closures, 32 VDC maximum, 1A noninductive load, for system alarm, range alert/or calibration alert, concentration alert/or calibration alert. Alarm signals are available on the RS485 interface. |
| Software Features: |
Displays ppmv or ppbv moisture reading or dew point, timer status, and instrument status |
| Environmentals: |
4º to 40ºC (40º to 105ºF), 90% relative humidity, non-condensing, noncorrosive atmosphere. Optimal performance in ppbv applications is achieved when the ambient temperature is maintained within ±2ºC. |
| Utility Requirements: |
85 to 265 volts, 47 to 63 Hz |
| Mounting Options: |
Available for stand-alone or 19" rack installation |
| Dimensions(WxHxD): |
43.2 x 13.2 x 38.1 cm (17 x 5.2 x 15) |
| Rack-mount Version: |
Same as above except height is 22.1 cm (8.71 in) |
| Net Weight: |
16.79 kg (37 lb) |
Product Data Sheet
Application Notes
- NEC Class I, DIV 2, Groups A,B,C,D
- EU EN 50081/82
- EN61010-1
|
|
 |
 |
|
 |