| Compatible Gases: |
Inerts (He, Ar, Ne, Xe, Kr) O2, H2, N2. Contact AMETEK to confirm compatibility with other gases. |
| Range: |
Calibrated from 0 to 150 ppbv. Trend indication to 1000 ppbv. |
| Limit of Detection: |
1.0 ppbv nominal |
| Accuracy: |
±1 ppbv or ±10% of the reading, whichever is greater |
| Sensitivity: |
Better than 1 ppbv |
| Response Time: |
Typically 80% of a 25 ppbv step change in either direction in 10 minutes or less |
| Inlet Pressure: |
138 to 345 kPa (20 to 50 psig). Specified performance is obtained when the inlet gas pressure is maintained within ±17 kPa (±2.5 psi) |
| Exhaust Pressure: |
Atmospheric |
| Sample Flow Requirement: |
<0.5 slpm |
| Inlet Gas Temperature: |
0º to 100ºC (32º to 212ºF). Optimal results are obtained when the inlet gas temperature is maintained at 60ºC (140ºF) |
| Outputs: |
Four line by 20 character LCD display
One self-powered 4 to 20 mA, into 100 to 500 ohm load analog output; can be configured for loop-powered operation
RS485 and RS232 serial ports |
| Alarms: |
System alarm, concentration alarm, and data valid 30 VAC or 60 VDC max, 50 VA or 1A max, resistive |
| Environmental Conditions: |
Ambient temperature range 10º to 30ºC (50º to 86ºF)
Optimal results are obtained when ambient is maintained within ±5ºC (±9ºF)
Relative humidity 90%, noncondensing
Pollution degree 2
Maximum altitude 2000 meters (6560 ft)
Installation category II
Indoor Use Only |
| Utility Requirements: |
100 to 132 VAC or 230 VAC ±10%, 47 to 63 Hz, 185W
Instrument Air: 550 to 690 kPa (80 to 100 psi)
-40ºC dew point |
| Mounting Configuration: |
19" rack |
| Dimensions(WxHxD): |
48x17.7x50.9 cm (19x7x20 in) |
| Net Weight: |
15.9 kg (35 lb) |