Model ASOMA 682T-HP Sulfur analyzer

Model ASOMA 682T-HP

The ASOMA 682T-HP is designed for the on-line analysis of sulfur in highly viscous hydrocarbons such as crude and residual oil using XRT (X-ray Transmission) technology. This system is ideal for high temperature and pressure sample streams. With its rugged, time-proven proportional counter detector and X-ray tube source, the 682T-HP gives superb analytical precision with a sulfur limit of detection of 200 ppm. The 682T-HP key features include a fully integrated densitometer, automatic density compensation, touch-screen controls, password protected and rapid update rates.

Overview
  • Ideal for analysis of crude and other highly viscous hydrocarbons, pipelines, terminals, blending
  • Sample stream temperatures in excess of 662˚F (350˚C)
  • Proprietary window in flowcell can withstand pressures up to 800psig/55 bar (consult factory for higher pressures)
  • Certified to industry standards
  • Large flowcell diameter (2 inches/5 cm) reduces interferences
  • Analysis range for sulfur of 0.02-6.0%

Operation of the analyzer is regulated by an integral processor. The processor monitors instrument parameters such as process temperature, density, and system diagnostics. It also provides a comprehensive suite of alarms and a standard platform for communicating to a plant-wide DCS. Due to its unique design and robust construction, the system has minimal, if any, sample conditioning and recovery requirements.

ASOMA Instruments was established in Austin, Texas, in 1979 as one of the first companies to offer low-cost benchtop ED-XRF instrumentation. Since then, the ASOMA brand has become an industry leader with the development and release of the Models 100, 200 and 200T-series, Phoenix, Phoenix II and Titan line of XRF analyzers, as well as instrumentation for measuring sulfur in hydrocarbon streams

  • Specifications +


    Process Conditions
    Normal Product Crude oil, heavy hydrocarbons, bunker fuels
    Sulfur Concentration 0.02 - 6%
    Pressure at flowcell Up to 800 psi *
    Flow rate at cell Up to 200 liters/ minute
    Temperature 662ºF/350ºC *
    Density Gauge Required (supplied by AMETEK or the customer)
    Ambient Temperatures Minimum 15ºC, maximum 40ºC *
       * Contact an AMETEK Representative for Special Requirements
    Enclosures
    Standard Stainless steel, NEMA 4X (IP-65), NEMA 7 Ex d
    Dimensions Controller 36x30x12 inch minimum
    Purge Type X for Class I, Division I, Groups A - D area classification, (ATEX Category 2) or Type Z for Class I, Division 2. Groups A - D area classification, (ATEX Category 3)
    Optional Heating and Air Conditioning System
    Power 115 or 230 VAC ± 10%, 50x60 Hertz, 80 W typical, 300 W max
    Controller
    Interface
    Class 1 DIV 2 External Touch Screen Interface, Power Switch, Valve Control Switch, Remote/ Local Control Switch, Capture Enable
    Class 1 DIV 1 Internal Touch Screen Interface, Power Switch, Valve Control Switch, Auto Start/Stop Switch, Capture Enable
    Standard Outputs 4-20 mA for sulfur and density, dry contact for common alarm, enclosure safety alarm, data valid, in bounds signal, RS-485 serial data
    Optional Outputs Profibus, Modbus, Ethernet
    Standard Inputs Densitometer input, dry contact for analyzer stop/start, mode switching
    Measurement Flowcell
    Technique X-ray Transmission (XRT)
    Analysis Time Nominally 100 seconds
    Typical One Sigma Precision ±50 ppm @ 0.1 Wt% sulfur
    ± 30 ppm @ 3.2 Wt% sulfur (consult AMETEK for application)
    X-ray Source X-ray tube
    X-ray Detector High resolution gas filled proportional counter Leak Detection: Capacitive leak sensor
  • Documents +

  • Sales & Service +


    Our ASOMA Service Center in Austin, TX is readily available to help with all your XRF needs for not only the ASOMA PHOENIX II and 682-HP, but also, for the ASOMA and SPECTRO legacy Models 200, 200T, 400, 600 On-Line series, Titan and PHOENIX.  Our ASOMA service center can help with:

    • Major Repairs
    • Routine Serivce and Maintenance
    • Applications Support
    • Basic Operation, Theory and Radiation Safety Training
    • Calibration
    • Leak Tests for isotope analyzers
    • Radiation Surveys
    • Analyzer Disposals
    • Radioisotope Source Replacements
    • Radioisotope Source Displosal
    • Application Studies
    • Sample Matrix Testing

    Ensuring your analyzer has the most recent electronic setup and software will better protect your investment.  For the most up to date software and hardware for you ASOMA products, be sure to contact the Austin Service Center for your repairs and questions.  As always, all phone and email support is offered at no charge.

    Free analyzer support includes:

    • Internet Diagnostics for the ASOMA PHOENIX II
    • Basic troubleshooting
    • Method and Calibration Support
    • Application and Feasibility Studies
    • Parts Questions

    We also have International service centers around the world to provide you with fast repairs and support wherever you are.

    For Service and Analyzer support contact us at:

    AMETEK Process Instruments
    ASOMA Service Center
    4903 West Sam Houston Parkway North, Suite A-400
    Houston, TX 77041

    +1 512-215-2064

    For analyzer orders and quotes please contact one of our Sales Specialists at:
    +1 412-828-9040
    +1 412-826-0399 (fax)

    For parts orders:
    +1 412-826-2441