Model 914 Continuous Emission Monitoring System

Model 914 Continuous Emission Monitoring System
Overview
The  Model 914 is a built-to-order Continuous Emission Monitor (CEM) that includes all the components needed for a compliance monitoring and reporting project. The system uses dry extractive sampling techniques designed to meet government regulations for all types of compliance monitoring.

The sample extraction system of the 914 combines a temperature controlled sample conditioning unit and a heated sampling line to ensure the integrity of the sample gas is reliably maintained. The sample gas is then dried (moisture removed) as required and introduced into the instruments for analysis.

AMETEK uses only instruments that have proven reliable over many years of use in its 914 systems. All of the CEM instruments are EPA certifiable. A recommended analyzer is the AMETEK Western Research Model 9000RM.

The instrument is mounted in a free-standing standard 19-inch open rack. Optional considerations are: an enclosed rack, a free-standing enclosure or a walk-in shelter for environmental reasons. A computer data acquisition system or programmable logic controller providing data collection and control functions for customer specific reporting and auto calibration is optional.

Other Products of Interest:

  • Specifications +


    Methodology Standard - NDUV, NDIR, Zirconium Oxide, Paramagnetic O2
    Calibration DAS/DCS/PLC or selected analyzer, initiated automatically using instrument air and calibration gas standards
    Sensitivity Instrument dependent
    Outputs Designed to meet customer specifications (standard voltage, current or serial)
    Range Designed to meet customer specifications
    Reporting Customer-defined format for reporting or regulatory agencies. EPA Part 60 and Part 75.
    Alarms/Calculations Designed to meet customer specifications
    Electrical Requirements 120/208 VAC or 110/220/240 VAC 50-60 Hz
    Typical Flow 2 to 10 SCFH
    Ambient Temperature 7 to 49 °C (45 to 120 °F), 5 to 95% relative humidity, non-condensing
    Physical Dimensions Application dependent
    Sample dryers – thermoelectric (Baldwin Environmental, Inc. and others)
    Current Configurations
    Analyzers Available
      Single species NDUV + paramagnetic oxygen Model 9900 SO2, H2S, NO, NO2, O2, NH3, Cl
      Multi species NDUV + paramagnetic oxygen Model 9900 for SO2, H2S, NO, NO2, NOx, O2, NH3, Cl
      Mass flow S-type pitot tube, optical, etc.
      Single species NDIR Various analyzer models available (Siemens, California Analytical, Servomex)
      Zirconium oxide Thermox O2 analyzers
    Probes Available AWR emission probe
    AWR mass emission probe
    M&C probe, c/w calibration port,
    (optional, blow back, and sample isolation valving)
    AWR velocity probe (used separate with M&C for mass)
    Universal probe for Division II applications
    Sample System Sample lines – temperature controlled, self-limiting electrical (O'Brien, Technical Heaters)
    Rack Components Basic 19” x 24” rack (no panels)
    Pressure control panel
    Temperature control panel
    Customer termination and system alarm panel
    Flow distribution panel(s)
    Rack fans (optional)
    Data Acquisition and Control System Allan Bradley PLC
    WTC and VIVICOM DAS (others available upon request)
    Shelter/Enclosure Open 19” x 24” rack
    Enclosed 19” x 24” rack
    Fiberglass or metal free-standing enclosure
    AMETEK’s standard walk-in shelters (optional desk/workbench)
    Area Classifications Model 914 CEM rack system in General Purpose and Division II (purged)
    Wetted Components 316 SS tubing, fittings (Swagelok only) and cell construction
    Teflon tubing
    Quartz glass (cell windows)
  • Documents +

  • Certifications +


    Model 914 CEM rack system is General Purpose and Division II (purged)