5820 Moisture Analyzer

Model 5812 Moisture Analyzer

Ideal for critical low-level moisture measurement applications

The 5820 is an easy-to-use process moisture analyzer that offers a wide range of performance features: exceptional accuracy, multi-gas compatibility, fast response speed, and a wide measurement range. Equipped with an online verification system, this state-of-the-art analyzer is designed to rapidly build and maintain operator confidence in its analyses.


Multi-gas compatibility
The 5820 combines excellent multi-gas compatibility with an easy-to-use operator interface. A single, simple menu selection is all that is needed to reconfigure the 5820 for a new gas type. The analyzer is compatible with virtually all non-corrosive gases including inert gases (He, Ar, Ne, Xe, Kr), H2, O2, and N2, air, and many specialty gases like sulfur hexafluoride.

Exceptional accuracy
With an accuracy of ±10 parts per billion by volume (ppbv) or ±10% of reading, the 5820 is perfect for moisture applications that require accurate results. Quartz crystal technology and an online verification system combine to constantly provide assurance that the analyzer is continuing to provide you with this superior level of performance.

Wide measurement range
The 5820 accurately measures from 0.01 parts per million by volume (ppmv) to 100 ppmv. While the recommended usable range is between 0.01 ppmv to 100 ppmv, the analyzer will provide measurements up to 1000 ppmv to capture the nature of a process upset.

Key Benefits:

  • Quartz crystal technology provides accuracy, speed, and calibration stability
  • Online verification system confirms analytical performance
  • Intuitive, easy-to-use interface with keypad and display allows access to operating variables
  • Multi-gas compatibility for periodic testing of multiple gas samples
  • Menu-driven gas selection eliminates manual adjustments


Other Products of Interest
3050 Moisture Analyzer Series
5100 TDLAS

  • Specifications +


    Compatible gases

    Inert gases (He, Ar, Ne, Xe, Kr), H2, O2, N2, air, some specialty gases such as sulfur hexafluoride. CO2 requires a custom measurement cell. (Contact the factory to confirm compatibility with other gases)

    Range  0 to 100 ppmv. Trend indication to 1000 ppmv. Display is software settable to show ppmv, ppbv, or dew point (requires pressure input)
    Reference dryer life Over 1,000,000 ppmv-hours nominal (e.g. dryer will last over five years with a 20 ppmv inlet moisture concentration)
    Limit of detection 10 ppb
    Accuracy  ±10 ppbv or ±10% of the reading, whichever is greater 
    Response time  63% of a step change in either direction in less than 5 minutes
    Inlet pressure  138 to 345 kPa (20 to 50 psig)
    Exhaust pressure  Atmospheric 
    Sample flow requirement Approximately 150 sccm in sample saver mode. Approximately 1 slpm with bypass open
    Sample gas temperature  0 to 100°C (32 to 212°F) 
    Outputs Four-line digital display
    Fully programmable 4-20 mA analog output
    RS485/RS232 serial communication
    Alarms Three dry contact closures, 32 VDC maximum, 1 A non-inductive load (moisture, data valid, and system alarm)
    Alarm signals are available on the RS485 interface
    Software features Displays ppmv or ppbv moisture reading or dew point, timer status, and instrument status
    Environmental conditions 5 to 50°C (41 to 122°F). 90% relative humidity, noncondensing, noncorrosive atmosphere
    Optimal performance in ppbv applications is achieved when the ambient temperature is maintained within ±2°C
    Utility requirements 85 to 265 volts, 47 to 63 Hz, 185 W Instrument air: 483 to 690 kPa (70 to 100 psi)
    Mounting configuration 19-inch rack/bench-top installation
    Dimensions (WxHxD) 483 x 133 x 419 mm (19 x 5.25 x 16.5 in.)
    Net Weight 12.7 kg (28 lb)
  • Documents +

  • Certifications +


    CE
    MET Certified to:

    • UL/CSA General Safety Requirements
    • NEC/CEC Class I, Div 2, Groups A, B, C, D, T4